Transcutaneous carbon dioxide monitoring.

作者信息

McLellan P A, Goldstein R S, Ramcharan V, Rebuck A S

出版信息

Am Rev Respir Dis. 1981 Aug;124(2):199-201. doi: 10.1164/arrd.1981.124.2.199.

Abstract

We studied changes in transcutaneous PCO2 in healthy volunteers and patients to establish the accuracy and practicability of a new cutaneous capnometer (HP47201A). In 9 healthy volunteers, as many as 4 experiments each were performed during each laboratory session. After cutaneous electrode stabilization, a gas mixture of CO2 in O2 was rebreathed under "open-loop" conditions allowing a rapid progressive increase in alveolar PCO2. The electrode stabilized on the skin with a time constant of 2.87 +/- 0.72 min. During rebreathing, alveolar PCO2 increased 4.25 +/- 0.63mm Hg/min. The simultaneous rate of increase in transcutaneous PCO2 was 2.91 +/- 0.59 mm Hg of CO2/min. Electrode response was also studied during steady-state conditions after a step increase in alveolar PCO2 of 10 mm Hg more than the resting end-tidal PCO2. The time constant of the skin electrode's response under these conditions was 3.2 +/- 1.9 min. We compared the arterial PCO2 of 25 blood samples with simultaneous transcutaneous PCO2 measurements in hemodynamically and metabolically stable patients. The range of arterial PCO2 values varied between 24 and 98 mm Hg, and the electrode remained stable during each 12-h study. Observed values closely approximated arterial PCO2 measurements, with transcutaneous readings 4 mm Hg more than measured arterial values (r = 0.98). We conclude that the cutaneous capnometer accurately and consistently reflects arterial PCO2 and has delay and response characteristics that may allow its use in clinical monitoring of patients.

摘要

我们研究了健康志愿者和患者经皮二氧化碳分压(PCO2)的变化,以确定一种新型皮肤二氧化碳监测仪(HP47201A)的准确性和实用性。在9名健康志愿者中,每次实验室 session 期间均进行多达4次实验。皮肤电极稳定后,在“开环”条件下重新呼吸氧气中的二氧化碳混合气体,使肺泡PCO2迅速逐渐升高。电极在皮肤上稳定下来的时间常数为2.87±0.72分钟。重新呼吸期间,肺泡PCO2以4.25±0.63mmHg/分钟的速度增加。经皮PCO2的同时增加速率为2.91±0.59mmHg的二氧化碳/分钟。在肺泡PCO2比静息呼气末PCO2升高10mmHg后,还在稳态条件下研究了电极反应。在这些条件下,皮肤电极反应的时间常数为3.2±1.9分钟。我们比较了25份血样的动脉PCO2与血流动力学和代谢稳定患者同时进行的经皮PCO2测量值。动脉PCO2值的范围在24至98mmHg之间,并且在每次12小时的研究期间电极保持稳定。观察值与动脉PCO2测量值非常接近,经皮读数比测量的动脉值高4mmHg(r = 0.98)。我们得出结论,皮肤二氧化碳监测仪准确且一致地反映动脉PCO2,并且具有延迟和反应特性,这可能使其可用于患者的临床监测。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验