Fogt E J, Dodd L M, Jenning E M, Clemens A H
Clin Chem. 1978 Aug;24(8):1366-72.
The Glucose-Controlled Insulin Infusion System (Biostator) is a modular, computerized, feedback control system for dynamic control of blood glucose concentrations in diabetics. This on-line glucose analyzer for use with whole blood utilizes a novel enzyme (glucose oxidase)-membrane configuration and an electrochemical cell to measure the H202 generated. The analyzer exhibits both short- and long-range stability, and instrument response and analyte concentration are linearly related over the full range of clinical interest. The response is fast, accurate, and precise, and permits determination of blood glucose within 2 min from the moment the blood leaves the patient. Correlation studies were completed to show the agreement between the Biostator Glucose Analyzer and the FDA's recommended hexokinase/glucose-6-phosphate dehydrogenase procedure on whole blood (e.g., average per cent recovered for 11 concentrations between 250 and 900 mg/liter was: hexokinase, 95.6%, Biostator Analyzer, 95.9%; bias and SDd, respectively, at low, normal, and high glucose values were: 12 and 41 mg/liter at the 500 mg/liter level; 4 and 52 mg/liter at the 1000 mg/liter level, and 4 and 128 mg/liter at the 4000 mg/liter level). No appreciable interference is observed with above-normal concentrations of bilirubin, uric acid, creatinine, sodium salicylate, or dextran. Platelet adhesion, which tends to decrease the useful life of the membrane, has been significantly decreased.
葡萄糖控制胰岛素输注系统(生物稳定器)是一种模块化、计算机化的反馈控制系统,用于动态控制糖尿病患者的血糖浓度。这种用于全血的在线葡萄糖分析仪采用了一种新型的酶(葡萄糖氧化酶)-膜结构和一个电化学池来测量产生的H202。该分析仪具有短期和长期稳定性,在整个临床关注范围内,仪器响应与分析物浓度呈线性关系。响应快速、准确且精确,能在血液离开患者后2分钟内测定血糖。完成了相关性研究,以显示生物稳定器葡萄糖分析仪与美国食品药品监督管理局推荐的全血己糖激酶/葡萄糖-6-磷酸脱氢酶法之间的一致性(例如,250至900毫克/升之间11种浓度的平均回收率为:己糖激酶法95.6%,生物稳定器分析仪95.9%;在低、正常和高血糖值时的偏差和标准差分别为:500毫克/升水平时为12和41毫克/升;1000毫克/升水平时为4和52毫克/升,4000毫克/升水平时为4和128毫克/升)。未观察到胆红素、尿酸、肌酐、水杨酸钠或右旋糖酐浓度高于正常水平时有明显干扰。血小板黏附现象(这往往会缩短膜的使用寿命)已显著减少。