Diaphragmatic EMG and transdiaphragmatic pressure measurements with a single catheter.

作者信息

Onal E, Lopata M, Ginzburg A S, O'Connor T D

出版信息

Am Rev Respir Dis. 1981 Nov;124(5):563-5. doi: 10.1164/arrd.1981.124.5.563.

Abstract

A single gastroesophageal catheter was used for simultaneous measurements of diaphragmatic electromyogram (EMGdi), esophageal, and gastric pressures in 10 normal volunteers. The catheter consisted of 2 polyethylene tubes, each with an outer diameter of 1.70 mm and an inner diameter of 1.19 mm, 2 platinum wire coils, and esophageal and gastric latex balloons. In all subjects studied, the increase in EMGdi, quantified as the average rate of rise of inspiratory moving average activity, and transdiaphragmatic pressure (Pdi) were linearly related to the increase in end-tidal PCO2 during CO2 rebreathing (range of r, 0.88 to 0.99 and 0.85 to 0.99, respectively). The relation between changes in EMGdi and that of Pdi, mean inspiratory flow, and occlusion pressure were also linear (range of r, 0.83-0.99, 0.73-0.99, and 0.84-0.99, respectively), indicating reliable recordings of EMGdi and Pdi during CO2 rebreathing in upright normal humans.

摘要

使用单个胃食管导管对10名正常志愿者同时进行膈肌肌电图(EMGdi)、食管和胃压力的测量。该导管由2根聚乙烯管组成,每根外径为1.70毫米,内径为1.19毫米,还有2个铂丝线圈以及食管和胃乳胶气囊。在所有研究对象中,EMGdi的增加(以吸气移动平均活动的平均上升速率量化)和跨膈压(Pdi)与二氧化碳重复呼吸期间潮气末PCO2的增加呈线性相关(r值范围分别为0.88至0.99和0.85至0.99)。EMGdi变化与Pdi、平均吸气流量和阻断压变化之间的关系也呈线性(r值范围分别为0.83 - 0.99、0.73 - 0.99和0.84 - 0.99),表明在直立正常人体进行二氧化碳重复呼吸期间,EMGdi和Pdi的记录可靠。