[麻醉气体和蒸汽对手术室的污染。II. 一氧化二氮的气相色谱分析]

[Contamination of the operating room by anesthetic gases and vapors. II. Gas chromatographic analysis of nitrous oxide].

作者信息

Cattaneo A D, Ferraiolo G, Rovatti M, Zattoni J, Donato A

出版信息

Minerva Anestesiol. 1981 Dec;47(12):827-30.

PMID:6801546
Abstract

The contamination by nitrous oxide of an operating room atmosphere was studied in a number of experiments, in the absence of personnel and using a gaschromatographic method. The evacuating device of the anesthesia machine proved to be ineffective to overcome the hazard of leaks in the breathing system, whereas the air conditioning flow rates (12 outside air changes per hour) minimized waste anesthetic gas concentrations.

摘要

在一系列实验中,在无人在场的情况下,采用气相色谱法对手术室空气中一氧化二氮的污染情况进行了研究。结果表明,麻醉机的抽气装置无法有效克服呼吸系统泄漏的危害,而空调换气率(每小时12次室外空气更换)可将麻醉废气浓度降至最低。