Suppr超能文献

监测引导的比浊法系统评估

Evaluation of a monitor guided nephelometric system.

作者信息

Wider G, Kulnigg E, Molinari E, Hotschek H, Bayer P M

出版信息

J Clin Chem Clin Biochem. 1982 Jan;20(1):1-8. doi: 10.1515/cclm.1982.20.1.1.

Abstract

The most common parameters in the specific protein field, namely the immunoglobulins G, A and M were investigated on the recently developed Immuno Video Nephelometer System (IVNS). This system consists of the nephelometer, a microprocessor controlled program and a monitor screen, where instructions, standard curve and results are displayed. Scattered light of immuno-complexes is measured at equilibrium after incubation of prediluted antigen-antibody mixtures. Data were compared with those obtained by radial immunodiffusion (RID), rate nephelometry (Beckman Immunochemistry System-ICS) and immunoturbidimetry (ENI-Gemsaec). Intrabatch variation on the Immuno Video Nephelometer System was found to be good (CV-2.6-3.7%) and day to day variation was satisfactory (CV-3.3-8.6%). There was also good correlation between the values found on the Immuno Video Nephelometer System and those of the other methods (correlation coefficient of 0.94-0.98). Instrumentation advantages, operation procedure and necessity of antigen excess check are discussed in detail.

摘要

在最近开发的免疫视频比浊仪系统(IVNS)上,对特定蛋白质领域中最常见的参数,即免疫球蛋白G、A和M进行了研究。该系统由比浊仪、微处理器控制程序和显示屏组成,显示屏上会显示操作说明、标准曲线和结果。在预稀释的抗原-抗体混合物孵育后达到平衡时,测量免疫复合物的散射光。将数据与通过放射免疫扩散(RID)、速率比浊法(贝克曼免疫化学系统-ICS)和免疫比浊法(ENI-Gemsaec)获得的数据进行比较。发现免疫视频比浊仪系统的批内变异良好(CV为2.6%-3.7%),日间变异也令人满意(CV为3.3%-8.6%)。免疫视频比浊仪系统测得的值与其他方法的值之间也具有良好的相关性(相关系数为0.94-0.98)。详细讨论了仪器优势、操作程序以及检查抗原过量的必要性。

相似文献

1
Evaluation of a monitor guided nephelometric system.监测引导的比浊法系统评估
J Clin Chem Clin Biochem. 1982 Jan;20(1):1-8. doi: 10.1515/cclm.1982.20.1.1.
2
Evaluation of the new Behring Nephelometer in the measurement of M proteins.
Am J Clin Pathol. 1989 Jul;92(1):93-6. doi: 10.1093/ajcp/92.1.93.
4
Survey of a new nephelometric system.
Ric Clin Lab. 1982 Jan-Mar;12(1):313-22. doi: 10.1007/BF02909338.
7
Evaluation of a Behring laser-nephelometer prototype in the measurement of IgG, IgA and IgM.
Clin Biochem. 1978 Feb;11(1):28-31. doi: 10.1016/s0009-9120(78)80040-x.
10
A comparative study of three approaches to the routine quantitation of human serum proteins.
Clin Biochem. 1978 Jun;11(3):112-6. doi: 10.1016/s0009-9120(78)90126-1.