Suppr超能文献

使用Technicon RA - 1000测定免疫球蛋白G、A和M。

Determinations of immunoglobulins G, A, and M in the Technicon RA-1000.

作者信息

Ng R H, Altaffer M, Statland B E

出版信息

Clin Chem. 1985 Sep;31(9):1554-7.

PMID:4028403
Abstract

We evaluated the Technicon RA-1000 "Random Access" analyzer for the measurements of immunoglobulins G, A, and M in serum by turbidimetry. For various concentrations of serum pools the total CV ranged from 3.0 to 4.5% for IgG, from 3.1 to 3.2% for IgA, and from 4.8 to 5.4% for IgM. The linearity of the standard curves was good in the clinically useful ranges. Results for patients' samples correlated well with those determined with the Beckman Auto ICS (rate nephelometric method). Normal reference intervals (based on data for 200 healthy blood donors) are 7.67 to 16.87 g/L for IgG, 0.87 to 3.89 g/L for IgA, and 0.61 to 2.70 g/L for IgM. The RA-1000 requires only 8 microL for all three tests and can analyze 25 specimens for all three proteins in 37 min. Compared with other methods for immunoglobulin assays, the RA-1000 has higher throughput and offers significant savings in labor and reagent costs.

摘要

我们评估了Technicon RA - 1000“随机存取”分析仪通过比浊法测定血清中免疫球蛋白G、A和M的性能。对于不同浓度的血清混合液,IgG的总变异系数(CV)范围为3.0%至4.5%,IgA为3.1%至3.2%,IgM为4.8%至5.4%。在临床有用范围内,标准曲线的线性良好。患者样本的检测结果与用贝克曼自动ICS(速率散射比浊法)测定的结果相关性良好。(基于200名健康献血者的数据)IgG的正常参考区间为7.67至16.87 g/L,IgA为0.87至3.89 g/L,IgM为0.61至2.70 g/L。RA - 1000的三项检测仅需8微升样本,且能在37分钟内分析25份标本的所有三种蛋白质。与其他免疫球蛋白检测方法相比,RA - 1000具有更高的通量,可显著节省劳动力和试剂成本。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验