Kobayashi Y, Yahata M, Watanabe F, Miyai K
J Steroid Biochem. 1982 Apr;16(4):521-4. doi: 10.1016/0022-4731(82)90073-5.
Direct solid phase fluoroimmunoassay of serum cortisol was established by using fluorescein isothiocyanate labelled anti-cortisol antibody and cortisol conjugated polyacrylamide beads. Sodium salicylate was used as a blocking agent for cortisol binding proteins. The sensitivity of this assay was 0.2 ng/assay tube and the measurable range was from 2 to 100 micrograms/dl using 10 microliters of serum. Intra- and inter-assay coefficients of variation were 11.6% (mean +/- SD 19.8 +/- 2.3 micrograms/dl, n = 5): and 12.1% (mean +/- SD 22.8 +/- 2.8 micrograms/dl, n = 5), respectively. THe accuracy was estimated from recovery study and the average recovery was 100.2%. Cortisol values determined by the present method correlated well with those determined by radioimmunoassay (r = 0.98, y = 0.98 x + 0.04, n = 35). The proposed assay satisfied the standard criteria of dilution, accuracy, and precision, and is applicable to routine measurement of serum cortisol.
采用异硫氰酸荧光素标记的抗皮质醇抗体和皮质醇偶联的聚丙烯酰胺珠建立了血清皮质醇直接固相荧光免疫分析法。水杨酸钠用作皮质醇结合蛋白的封闭剂。该分析方法的灵敏度为0.2 ng/测定管,使用10微升血清时可测量范围为2至100微克/分升。批内和批间变异系数分别为11.6%(平均值±标准差19.8±2.3微克/分升,n = 5)和12.1%(平均值±标准差22.8±2.8微克/分升,n = 5)。通过回收率研究评估准确性,平均回收率为100.2%。用本法测定的皮质醇值与放射免疫分析法测定的值相关性良好(r = 0.98,y = 0.98x + 0.04,n = 35)。所提出的分析方法符合稀释、准确性和精密度的标准准则,适用于血清皮质醇的常规测定。