犬经食管多普勒检测空气栓塞的评估。

Evaluation of transesophageal Doppler detection of air embolism in dogs.

作者信息

Martin R W, Colley P S

出版信息

Anesthesiology. 1983 Feb;58(2):117-23. doi: 10.1097/00000542-198302000-00003.

Abstract

The authors assessed the use of a cylindrical sensor which transceives (transmits and receives) ultrasound in a 360 degree arc mounted on a standard esophageal stethoscope catheter to detect air embolism in dogs. They used electronic circuitry developed specifically to provide a continuous analog recording of high frequency Doppler energy as well as an audible signal. They found that the esophageal Doppler sensor was easy to position. In 25 of 30 dogs, the system distinguished Doppler sounds of venous air emboli that were present either in the superior vena cava, right atrium, right ventricle, or pulmonary artery from normal cardiovascular sounds. In the remaining five animals, Doppler heart and air emboli sounds were initially of poor quality, but improved with aspiration of air from the esophagus. In another five dogs, arterial air emboli produced by left ventricular injections also were detected. The esophageal sensor was sensitive to both venous and arterial air emboli ranging from 0.05--0.2 ml of air, and the authors consistently detected repeated intravenous injections of air throughout a 5-h time period. Optimal position of the sensor for detection of venous air emboli was at the level of the superior vena cava above its junction with the right atrium. Optimal position to detect arterial air emboli introduced via a left ventricular catheter was at the level of the aortic arch. Tissue analysis of the esophagus revealed no morphologic damage due to the sensor or transmitted energy.

摘要

作者评估了一种圆柱形传感器的用途,该传感器安装在标准食管听诊器导管上,能以360度弧形收发超声波,用于检测犬类的空气栓塞。他们使用了专门开发的电子电路,以提供高频多普勒能量的连续模拟记录以及可听信号。他们发现食管多普勒传感器易于定位。在30只犬中的25只中,该系统能够将存在于上腔静脉、右心房、右心室或肺动脉中的静脉空气栓塞的多普勒声音与正常心血管声音区分开来。在其余5只动物中,多普勒心脏和空气栓塞声音最初质量较差,但通过从食管中抽出空气后有所改善。在另外5只犬中,还检测到了由左心室注射产生的动脉空气栓塞。食管传感器对0.05 - 0.2毫升空气范围内的静脉和动脉空气栓塞均敏感,作者在5小时的时间段内持续检测到了重复静脉注射空气的情况。检测静脉空气栓塞的传感器最佳位置是在上腔静脉与右心房交界处上方的水平。检测通过左心室导管引入的动脉空气栓塞的最佳位置是在主动脉弓水平。对食管的组织分析显示,传感器或传输能量未造成形态学损伤。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验