使用稳定温度平台炉和塞曼背景校正直接测定尿液中的镉。

Direct determination of cadmium in urine with use of a stabilized temperature platform furnace and Zeeman background correction.

作者信息

Pruszkowska E, Carnrick G R, Slavin W

出版信息

Clin Chem. 1983 Mar;29(3):477-80.

PMID:6825258
Abstract

In this determination of Cd in urine, simple aqueous standards were used to which NaCl and the matrix modifier were added. The urine was diluted fivefold with water. The mean analytical recovery of added Cd for urine samples was 101%, with individual variations of less than 4%. We used the stabilized temperature platform furnace, Zeeman background correction, pyrolytically coated graphite tubes, and (NH4)2HPO4 plus HNO3 as a matrix modifier. The sensitivity of the method provided a characteristic amount of 0.35 pg of Cd per 0.0044 A X s, obtained with integrated absorbance readings. The absolute Cd detection limit in urine was 0.15 pg, corresponding to 0.04 microgram/L of urine. Lower relative detection limits for Cd in urine can be attained if the analytical situation demands it.

摘要

在本次尿样中镉的测定中,使用了添加了氯化钠和基体改进剂的简单水溶液标准品。尿样用水稀释了五倍。尿样中添加镉的平均分析回收率为101%,个体差异小于4%。我们使用了稳定温度平台炉、塞曼背景校正、热解涂层石墨管以及(NH4)2HPO4加HNO3作为基体改进剂。该方法的灵敏度在积分吸光度读数下,每0.0044 A×s可获得0.35 pg镉的特征量。尿样中镉的绝对检测限为0.15 pg,相当于尿样中0.04微克/升。如果分析情况需要,可实现更低的尿样中镉的相对检测限。