Palm R, Sjöström R, Hallmans G
Clin Chem. 1983 Mar;29(3):486-91.
This method for direct determination of Zn in cerebrospinal fluid (CSF) involves flame atomic absorption spectrophotometry with a pulse nebulizer technique. Standard solutions of Zn in 150 mmol/L NaCl were used. We could account for 88% of added standard with the method in individual samples from 10 patients and in pooled CSF. The method is acceptably precise, CVs in pooled CSF ranging from 4 to 12%. The mean CSF-Zn concentration for nine healthy men was 0.18 (SD 0.04) mumol/L and for nine healthy women 0.15 (SD 0.03) mumol/L, a statistically insignificant difference. These values are lower than those in previous reports, which may have been the result of contamination problems, nonatomic absorption, or nonstandardized sampling. In the healthy volunteers, the CSF-Zn concentration was positively correlated with serum-Zn, CSF-protein, and CSF-albumin concentrations, as well as with the CSF/serum ratio for albumin.
这种直接测定脑脊液(CSF)中锌(Zn)的方法采用带有脉冲雾化器技术的火焰原子吸收分光光度法。使用了150 mmol/L氯化钠中的锌标准溶液。用该方法可在来自10名患者的单个样本和合并脑脊液中测定88%添加的标准物质。该方法精密度可接受,合并脑脊液中的变异系数(CV)为4%至12%。9名健康男性的脑脊液锌平均浓度为0.18(标准差0.04)μmol/L,9名健康女性为0.15(标准差0.03)μmol/L,差异无统计学意义。这些值低于先前报告中的值,这可能是污染问题、非原子吸收或采样未标准化所致。在健康志愿者中,脑脊液锌浓度与血清锌、脑脊液蛋白、脑脊液白蛋白浓度以及白蛋白的脑脊液/血清比值呈正相关。