Pavlicek W, Geisinger M, Castle L, Borkowski G P, Meaney T F, Bream B L, Gallagher J H
Radiology. 1983 Apr;147(1):149-53. doi: 10.1148/radiology.147.1.6828720.
The effect of nuclear magnetic resonance (NMR) imaging on six representative cardiac pacemakers was studied. The results indicate that the threshold for initiating the asynchronous mode of a pacemaker is 17 gauss. Radiofrequency levels are present in an NMR unit and may confuse or possibly inhibit demand pacemakers, although sensing circuitry is normally provided with electromagnetic interference discrimination. Time-varying magnetic fields can generate pulse amplitudes and frequencies to mimic cardiac activity. A serious limitation in the possibility of imaging a patient with a pacemaker would be the alteration of normal pulsing parameters due to time-varying magnetic fields.
研究了核磁共振(NMR)成像对六种代表性心脏起搏器的影响。结果表明,启动起搏器异步模式的阈值为17高斯。NMR设备中存在射频水平,可能会使按需型起搏器产生混淆或可能对其产生抑制作用,尽管传感电路通常具备电磁干扰辨别能力。随时间变化的磁场可产生脉冲幅度和频率以模拟心脏活动。由于随时间变化的磁场导致正常脉冲参数改变,这将严重限制对佩戴起搏器患者进行成像的可能性。