Suppr超能文献

采用落射照明和透照法对红细胞速度进行体外和体内测量。

In vitro and in vivo measurement of red cell velocity with epi- and transillumination.

作者信息

Harper S L, Bohlen H G

出版信息

Microvasc Res. 1983 Mar;25(2):186-93. doi: 10.1016/0026-2862(83)90014-6.

Abstract

Measurement of red cell velocity with the dual-slit cross-correlation method in glass capillary tubes during transillumination indicates that the measured velocity must be divided by a correction factor of approximately 1.6 to equal the average velocity calculated from a known flow and inner diameter. Whether the same correction factor exists when red cell velocity is measured during epiillumination is questionable. Red cell velocity was measured with the dual-slit correlation method nearly simultaneously using epi- (EL) and transillumination (TL) while glass tubes (40-100 microns, i.d.) were pump perfused with whole human blood (hematocrit 39-42%). With TL, the measured velocity is 1.58 +/- 0.07 (SEM) times the calculated average velocity, whereas a factor of 2.04 +/- 0.04 (SEM) was obtained with epiillumination. When intestinal arterioles with approximately the same inner diameters and flow velocities as the glass tubes were used, the ratio of velocities measured with TL to EL was 1.21 +/- 0.02 (SEM) as compared to 1.31 +/- 0.09 (SEM) for glass tubes using TL and EL of the tube at the same pump flow. This similarity of TL to EL velocity ratios for glass tubes and microvessels may be fortuitous or indicate that comparable flow properties and measurement conditions exist for in vitro and in vivo situations. The major finding of the study is, however, that different velocity correction factors exist for EL and TL measurements when the dual-slit correlation method is used to estimate red cell velocities in tubes of an internal diameter of 40-100 microns at normal hematocrits.

摘要

在透照期间用双缝互相关方法测量玻璃毛细管中红细胞速度表明,测得的速度必须除以约1.6的校正因子才能等于根据已知流量和内径计算出的平均速度。在落射照明期间测量红细胞速度时是否存在相同的校正因子尚不确定。在用双缝相关方法几乎同时使用落射照明(EL)和透照(TL)测量红细胞速度的同时,用全血(血细胞比容39 - 42%)对内径为40 - 100微米的玻璃管进行泵灌注。对于透照,测得的速度是计算出的平均速度的1.58±0.07(标准误)倍,而落射照明时得到的因子为2.04±0.04(标准误)。当使用内径和流速与玻璃管大致相同的肠小动脉时,与在相同泵流量下使用玻璃管的透照和落射照明相比,透照与落射照明测量的速度比为1.21±0.02(标准误),而玻璃管的该比值为1.31±0.09(标准误)。玻璃管和微血管的透照与落射照明速度比的这种相似性可能是偶然的,或者表明体外和体内情况存在可比的流动特性和测量条件。然而,该研究的主要发现是,当使用双缝相关方法在正常血细胞比容下估计内径为40 - 100微米的管中的红细胞速度时,落射照明和透照测量存在不同的速度校正因子。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验