Fleming W H, Sarafian L B, Yarbrough J W, Kugler J D, Mooring P K
Ann Thorac Surg. 1983 May;35(5):535-7. doi: 10.1016/s0003-4975(10)60428-6.
Three children with a pacemaker sustained similar lead fractures within a two-year period. Each fracture occurred in the corkscrew portion of a Medtronic Model 6917 epicardial lead. All patients were boys. The leads had been in place for 12, 45, and 43 months prior to fracture. None of the patients had sustained major trauma. Fracture of the corkscrew lead has been considered an uncommon complication. However, three lead fractures in the corkscrew area in a population of 60 children followed at this institution indicate that this may be a more common cause of pacemaker-system malfunction than previous data suggest. A plea is made for reporting all pacing failure to the manufacturer.
三名植入起搏器的儿童在两年内发生了类似的导线骨折。每例骨折均发生在美敦力6917型心外膜导线的螺旋部分。所有患者均为男孩。骨折前导线分别已植入12个月、45个月和43个月。所有患者均未遭受重大创伤。螺旋导线骨折一直被认为是一种罕见的并发症。然而,在本机构随访的60名儿童中,有3例在螺旋区域发生导线骨折,这表明这可能是起搏器系统故障比以往数据显示的更为常见的原因。呼吁向制造商报告所有起搏故障情况。