猫升主动脉中速度、直径、血流量和压力的同步测量。一种用于连续监测主动脉血流量的可提取多普勒超声探头的评估。

Simultaneous measurement of velocity, diameter, flow and pressure in the ascending aorta of cats. Evaluation of an extractable Doppler ultrasound probe for continuous monitoring of aortic blood flow.

作者信息

Matre K, Segadal L

机构信息

Department of Surgery, University of Bergen, Haukeland Hospital, Norway.

出版信息

Cardiovasc Res. 1988 Dec;22(12):855-63. doi: 10.1093/cvr/22.12.855.

Abstract

To evaluate an extractable Doppler ultrasound probe designed for monitoring postoperative aortic flow in infants, simultaneous measurements of velocity, diameter, electromagnetic flow estimates, and pressure in the ascending aorta were recorded in cats. Doppler flow calculators were designed to provide flow estimates based on preset constant or measured instantaneous aortic diameter. Doppler sample volume was extended to about 1.2 X 1.6 X 4.0 mm. By pharmacological interventions, a wide range of flow rates and pressures were obtained, some of which were outside physiological values. The prototype probe described here was well suited for implantation in aortas with outer diameter 8-12 mm. Correlation coefficients between Doppler and electromagnetic estimated flow for 140 samples were 0.908 and 0.959 for constant and instantaneous diameter, respectively. Mean difference between the two methods was calculated, as well as the 95% confidence limits (2SD) of the differences. Limits of agreement, in stable haemodynamic situations without drugs, were -20.0 and 22.4% of mean average flow for constant diameter Doppler flow estimates, and -14.5 and 15.1% of electromagnetic flow values for instantaneous diameter estimates. These limits increased when all samples were taken into account, but the limits for instantaneous diameter estimates were reduced from -33.1 and 32.4 to -23.9 and 19.3% after removal of data obtained during isoprenaline infusion, which introduced extreme high flow/low pressure situations. This Doppler method seems well suited for postoperative monitoring of aortic flow in infants. Errors observed in larger aortas with an asymmetric flow profile do not seem to be present in smaller vessels within normal values of flow and pressure.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

为评估一款专为监测婴儿术后主动脉血流而设计的可提取式多普勒超声探头,研究人员在猫身上同步记录了升主动脉的速度、直径、电磁血流估计值和压力。多普勒血流计算器旨在根据预设常数或测量的瞬时主动脉直径提供血流估计值。多普勒取样容积扩展至约1.2×1.6×4.0毫米。通过药物干预,获得了广泛的流速和压力范围,其中一些超出了生理值。此处描述的原型探头非常适合植入外径为8 - 12毫米的主动脉。140个样本的多普勒估计血流与电磁估计血流之间的相关系数,对于常数直径和瞬时直径分别为0.908和0.959。计算了两种方法之间的平均差异以及差异的95%置信限(2SD)。在无药物的稳定血流动力学情况下,对于常数直径多普勒血流估计,一致性界限为平均平均血流的 - 20.0%和22.4%,对于瞬时直径估计为电磁血流值的 - 14.5%和15.1%。当考虑所有样本时,这些界限增加,但在去除异丙肾上腺素输注期间获得的数据(该数据引入了极高流量/低压情况)后,瞬时直径估计的界限从 - 33.1%和32.4%降至 - 23.9%和19.3%。这种多普勒方法似乎非常适合婴儿术后主动脉血流的监测。在具有不对称血流分布的较大主动脉中观察到的误差,在血流和压力正常范围内的较小血管中似乎不存在。(摘要截断于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验