Imhof H, Doi K
Radiology. 1978 Oct;129(1):173-8. doi: 10.1148/129.1.173.
2X magnification employing a 200-mu focal spot and an Alpha 8-XM screen/film system was applied to oral cholecystography and the results compared with those for the conventional contact technique with the Par-RP system. The basic imaging properties of the system, as well as phantom studies, indicated that the image quality obtained with magnification is comparable to or better than that for the conventional technique. In clinical studies on the detection of gallstones, the conventional technique revealed 5 true-positive and 17 true-negative cases and 1 false-positive and 2 false-negative cases, while the magnification technique provided 7 true-positive and 18 true-negative cases but no false cases. With the magnification technique the skin dose was reduced to approximately half that for the conventional contact technique.
采用200微米焦点和Alpha 8-XM屏/片系统进行2倍放大用于口服胆囊造影,并将结果与使用Par-RP系统的传统接触技术的结果进行比较。该系统的基本成像特性以及体模研究表明,放大成像获得的图像质量与传统技术相当或更好。在胆结石检测的临床研究中,传统技术显示5例假阳性和17例假阴性病例以及1例假阳性和2例假阴性病例,而放大技术提供了7例假阳性和18例假阴性病例但无假病例。使用放大技术时,皮肤剂量降低至传统接触技术的约一半。