Suppr超能文献

[对放射诊断学中放大技术的批判性评论(作者译)]

[Critical comment on enlargement techniques in diagnostic radiology (author's transl)].

作者信息

Schmidt-Hieber M

出版信息

Rofo. 1975 Aug;123(2):175-9. doi: 10.1055/s-0029-1230174.

Abstract

Detail sharpness in radiographic magnification depends upon the properties of the focal spot, the screen and upon object motion. With direct magnification the finite size of the focal spot causes an increased penumbra. Uneven distribution of focal spot radiation leads to formation of false images. Radiographic magnification is rational when using a minute focal spot in combination with a poorly resolving (high speed) screen while the object remains stationary. Using high detail screens or non screen films there is no improved but possibly inferior resolution to non-magnified images, even though 0.1 mm x 0.1 mm foci are employed. Radiographic magnification performed with focal spots larger than 0.3 mm x 0.3 mm is not reasonable. Magnification procedures cause a considerable radiation exposure to the patient.

摘要

X线放大摄影中的细节清晰度取决于焦点、增感屏的特性以及物体的移动。在直接放大摄影中,焦点的有限尺寸会导致半影增大。焦点辐射的不均匀分布会导致伪像的形成。当使用微小焦点与分辨率低(高速)的增感屏相结合且物体保持静止时,X线放大摄影是合理的。即使使用0.1mm×0.1mm的焦点,使用高清晰度增感屏或无屏胶片时,分辨率并不比未放大的图像提高,甚至可能更低。使用大于0.3mm×0.3mm的焦点进行X线放大摄影是不合理的。放大摄影程序会使患者受到相当大的辐射暴露。