Raphael E, Sandrik J L, Klapper L
Am J Orthod. 1981 Aug;80(2):136-44. doi: 10.1016/0002-9416(81)90214-1.
The degree of angular rotation of rectangular orthodontic wires in rectangular molar tubes was measured using a rotatable mechanical stage on a Unitron metallograph and compared to theoretical data. It was found that square 0.016 inch wire did not bind in 0.018 by 0.022 inch or larger mandrel- or inconel-formed tubes, whereas rotations of 11 to 23 degrees were noted in cast tubes of this dimension. Although 0.016 by 0.022 inch wire did bind in these tubes, the degree of rotation was far greater than expected on the basis of theoretical calculations or clinical expectations. The results indicated that it was unrealistic to expect fine control of a few degrees of torque to occur as expected for delicate clinical control.