五种手动凝血分析仪的比较。

A comparison of five manually operated coagulation instruments.

作者信息

Beckala H R, Leavelle D E, Didisheim P

出版信息

Am J Clin Pathol. 1978 Jul;70(1):71-5. doi: 10.1093/ajcp/70.1.71.

Abstract

The results of activated partial thromboplastin time (APTT) and prothrombin time (PT) determinations on the single-channel Coag-a-Mate, Coagulyzer Jr., Clotek, Electra 650, and Fibrometer instruments were compared with regard to precision, day-to-day reproducibility of results, and accuracy. There was no significant difference among the machines when PT was analyzed. In the APTT study, the single-channel Coag-a-Mate had the best overall precision (coefficient of variation 2 to 4%), while the Coag-a-Mate and the Electra 650 had the best day-to-day reproducibility (coefficient of variation 2 to 5%). The comparison study indicated good correlation (r greater than or equal to 0.95) between the reference instrument (Bio/Data CA-15) and the single-channel Coag-a-Mate, Electra 650, and Clotek. The correlation for the Coagulyzer Jr. was also quite good (r = 0.95), but it could not detect some of the clots detected by the other machines. The Fibrometer showed significantly poorer correlation (r = 0.87) when compared with the other instruments.

摘要

对单通道凝血仪Coag-a-Mate、Coagulyzer Jr.、Clotek、Electra 650和纤维蛋白原测定仪Fibrometer的活化部分凝血活酶时间(APTT)和凝血酶原时间(PT)测定结果在精密度、结果的日常重复性和准确性方面进行了比较。分析PT时,各仪器之间无显著差异。在APTT研究中,单通道Coag-a-Mate总体精密度最佳(变异系数为2%至4%),而Coag-a-Mate和Electra 650的日常重复性最佳(变异系数为2%至5%)。比较研究表明,参考仪器(Bio/Data CA-15)与单通道Coag-a-Mate、Electra 650和Clotek之间具有良好的相关性(r≥0.95)。Coagulyzer Jr.的相关性也相当好(r = 0.95),但它无法检测到其他仪器检测到的一些凝块。与其他仪器相比,Fibrometer的相关性明显较差(r = 0.87)。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验