[Determination of the prothrombin time and of the activated cephalin time wit semi-automatic Electra 600 apparatus (author's transl)].

作者信息

Cazalet C, Jaffray P, Raichvarg D, Bonnaire Y, Guest M

出版信息

Ann Biol Clin (Paris). 1981;39(2):85-8.

PMID:7258752
Abstract

This work studies the quality of the response of the Electra 600 for the determination of the prothrombin time (PT) and the activated cephalin time (ACT). The intra-serial precision was good for the PT and the ACT with normal or pathological plasmas (CV 2%). The precision was excellent for the PT (r = 0.98; Electra 600-fibrometer). The correlation coefficient varies between 0.85 and 0.95 for the ACT depending on the nature of the activator chosen. Lactescent plasmas having a protein concentration greater than 85 g/l should be treated manually for measurement of the PT. No interference was noted with other biological substances: eg glucose, bilirubin, hemoglobin. The increasing addition of heparin indicates a correct sensitivity of the response with Electra 600 both for short and long ACT. Autoanalysis renders the determinations of PT and ACT independant of the manipulator, that of the PT being much more rapid than with the manual method

摘要

本研究旨在探讨Electra 600在测定凝血酶原时间(PT)和活化部分凝血活酶时间(ACT)时的反应质量。对于正常或病理性血浆,PT和ACT的批内精密度良好(变异系数CV为2%)。PT的精密度极佳(r = 0.98;Electra 600 - 纤维蛋白原测定仪)。根据所选激活剂的性质,ACT的相关系数在0.85至0.95之间变化。蛋白质浓度大于85 g/l的乳状血浆在测定PT时应手动处理。未发现与其他生物物质(如葡萄糖、胆红素、血红蛋白)存在干扰。肝素添加量的增加表明Electra 600对于短ACT和长ACT的反应具有正确的敏感性。自动分析使PT和ACT的测定独立于操作人员,PT的测定比手动方法快得多。