Wells W J, Stiles Q R
Ann Thorac Surg. 1981 Jan;31(1):86-9. doi: 10.1016/s0003-4975(10)61324-0.
A massive bolus of oxygen was forced into the right atrium and venous system of a patient during cardiopulmonary bypass using a Bentley BOS-10 oxygenator. The unique design of the oxygenator canister provides an explanation for the event. Methods are outlined to avoid this type of venous embolism.
在使用本特利BOS - 10氧合器进行体外循环期间,大量氧气团被强行注入一名患者的右心房和静脉系统。氧合器罐的独特设计为该事件提供了解释。文中概述了避免此类静脉栓塞的方法。