Comparison of 180 degrees and 360 degrees data collection in thallium-20 1 imaging using single-photon emission computerized tomography (SPECT): concise communication.

作者信息

Coleman R E, Jaszczak R J, Cobb F R

出版信息

J Nucl Med. 1982 Aug;23(8):655-60.

PMID:6980969
Abstract

Thallium-201 imaging using SPECT is being done with 180 degrees (RAO to LPO) data collection in some centers with single-gamma camera systems. Using our SPECT system with two gamma cameras, we have compared the effects of 180 degrees data collection without attenuation correction against 360 degrees collection with attenuation correction, using phantoms and patients. With a heart phantom in a chest phantom, TI-201 activities simulating "normal myocardium," "ischemia," "infarction," and "background" were placed in object contrast ratios (with respect to background) of 5.0, 2.0, and -1.0, respectively. The 180 degrees data gave image contrast ratios of 1.6, 0.2, and -0.8, and the 260 degrees data gave ratios of 1.5, 0.8, and -0.3, respectively. Uniform activity throughout the heart gave similar image contrast with both data-collection methods, but there was more variability with the 180 degrees collection than with 360 degrees collection. Since attenuation correction is available with the 260 degrees collection, the effects of attenuation are seen only on the 180 degrees collection images. In eight patients the image contrasts from the 180 degrees and 260 degrees collections are similar. For our two-camera SPECT system, the 360 degrees collection permits attenuation correction, has less variability in counting statistics, and gives contrast ratios like those of 180 degrees collection.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验