Drow D L, Manning D D
J Clin Microbiol. 1980 Jun;11(6):641-5. doi: 10.1128/jcm.11.6.641-645.1980.
An indirect sandwich enzyme-linked immunosorbent assay for the detection of the polysaccharide antigen of type 3 pneumococcus (SSS-III) is reported. Polystyrene balls with attached anti-SSS-III antibody serve as the solid phase, and antigen is detected using an alkaline phosphatase-labeled antiglobulin conjugate. The reaction is quantitated by spectrophotometry. Concentrations of antigen are estimated by comparison with standard curves prepared with purified SSS-III. For this assay, the practical lower limit of detection of SSS-III is approximately 2 to 3 ng/ml, thus making the test sensitivity about 25 times that reported for counterimmunoelectrophoresis. In preliminary tests with simulated human clinical specimens, none of the body fluids tested (spinal fluid, serum, urine, and sputum) interfered with detection of antigen, nor did they produce false-positive or false-negative results. Streptococcus pneumoniae type 3 whole organisms were readily detected in simulated clinically positive sputum specimens. Cross-reactions were not observed with Haemophilus influenzae type b, group B Streptococcus, Klebsiella pneumoniae, Escherichia coli, Staphylococcus aureus, and Pseudomonas aeruginosa.
本文报道了一种用于检测3型肺炎球菌多糖抗原(SSS-III)的间接夹心酶联免疫吸附测定法。附着有抗SSS-III抗体的聚苯乙烯球作为固相,使用碱性磷酸酶标记的抗球蛋白缀合物检测抗原。通过分光光度法定量反应。通过与用纯化的SSS-III制备的标准曲线比较来估计抗原浓度。对于该测定法,SSS-III的实际检测下限约为2至3 ng/ml,因此该测试的灵敏度约为对流免疫电泳报道灵敏度的25倍。在对模拟人类临床标本的初步测试中,所检测的体液(脑脊液、血清、尿液和痰液)均未干扰抗原检测,也未产生假阳性或假阴性结果。在模拟临床阳性痰液标本中很容易检测到3型肺炎链球菌全菌。未观察到与b型流感嗜血杆菌、B族链球菌、肺炎克雷伯菌、大肠杆菌、金黄色葡萄球菌和铜绿假单胞菌的交叉反应。