Graniewski-Wijnands H, Greve E L, Bakker D, Moed J L
Doc Ophthalmol. 1982 Dec 1;53(4):303-10. doi: 10.1007/BF00216791.
The Competer automatic campimeter was tested in 98 patients. The results of 31 visual fields with glaucomatous defects and 11 (para) central defects due to maculopathies are reported. The examination with the Competer will miss a certain amount of early glaucomatous visual field defects. The percentage depends on the selection of the visual fields. Any substantial glaucomatous visual field defect will be detected. The indication of typical configuration is not ideal. The examination with the Competer is not suitable for the detection of small (para) central defects, due to lack of stimulus-positions in the (para) central visual field. The Competer's meridional static perimetry device was not used in this study. The Competer has a large dynamic range and a good detection strategy. The stimulus distribution over the visual field could be improved.
对98例患者使用了Competer自动视野计进行检测。报告了31例青光眼性缺损视野及11例因黄斑病变导致的(旁)中心缺损视野的检测结果。使用Competer进行检测会遗漏一定数量的早期青光眼性视野缺损。遗漏百分比取决于视野的选择。任何明显的青光眼性视野缺损都会被检测到。典型形态的指示并不理想。由于(旁)中心视野缺乏刺激位置,使用Competer进行检测不适合检测小的(旁)中心缺损。本研究未使用Competer的子午线静态视野计装置。Competer具有较大的动态范围和良好的检测策略。视野上的刺激分布可以得到改善。