Broms P, Ivarsson A, Jonson B
Acta Otolaryngol. 1982 May-Jun;93(5-6):455-60. doi: 10.3109/00016488209130904.
A rhinomanometer for clinical work and research is presented. It adheres to the principle that pressure and flow should be recorded X-Y wise. A storage oscilloscope is used and the tracing pencilled down on a preprinted diagram via an oscillotracer. Anterior and posterior rhinomanometry can be performed. In the former variety, one nostril is occluded and connected to the pressure transducer by an adhesive tape. This does not deform or irritate the nasal airway. A mask for anaesthesia connects the nose to the pneumotachograph. Miniature transducers mounted on the pneumotachograph provide mechanical advantages. Simple means that allow dynamic calibration of the rhinomanometer are described. The cost to set up and operate the equipment is low.
介绍了一种用于临床工作和研究的鼻阻力计。它遵循压力和流量应按X-Y方式记录的原则。使用存储示波器,并通过示波器记录器将描记图铅笔绘制在预印图表上。可进行前鼻阻力测量和后鼻阻力测量。在前一种类型中,一个鼻孔被堵塞,并用胶带连接到压力传感器。这不会使鼻气道变形或产生刺激。用于麻醉的面罩将鼻子连接到呼吸流速计。安装在呼吸流速计上的微型传感器具有机械优势。描述了对鼻阻力计进行动态校准的简单方法。设置和操作该设备的成本较低。