Suppr超能文献

两种市售鼻阻力计的序贯临床比较。

Sequential clinical comparison between two commercially available rhinomanometers.

作者信息

Borges Dinis P, Guerra J B, Gomes A

机构信息

Department of Otorhinolaryngology, Hospital Pulido Valente, Lisbon, Portugal.

出版信息

Acta Otolaryngol. 1994 Jul;114(4):435-42. doi: 10.3109/00016489409126083.

Abstract

To assess nasal resistance in the treated as well as decongested states, two custom-built rhinomanometers (Rhino-Comp, manufactured by Cintec, Sweden, and Rhinomanometer 200, manufactured by Atmos, Germany) were sequentially employed in 53 non-selected consecutive patients referred for chronic or recurring nasal obstruction complaints, using the same active anterior rhinomanometric method. Final mean results from both equipments were submitted to paired statistical analysis. No correlation was found between the R150 results provided by the two rhinomanometers, albeit a 1.26 to 1.30 difference was encountered in the Rhinomanometer 200/Rhino-Comp ratio when the averaged mean values of all R150 results from both equipments were compared, suggesting that the discrepancy could be due to technical specificities intrinsic to each apparatus, with a tendency to produce systematic R150 differences. When the graphic aspect of the curves from both equipments were compared using the Broms system parameter V2 values, a remarkable correlation was seen between Rhino-Comp and Rhinomanometer 200 in expiratory (but not inspiratory) results, suggesting that the expiratory curves produced by the two equipments are graphically similar, at least until the Broms system radius 2 interception. Enough data was also found to support the possibility that the manufacturer's option to use a nose adaptor instead of adhesive tape for anterior nares occlusion with the Rhinomanometer 200 equipment could interfere with nasal vestibular function and thus artificially affect resistance measurements. Finally, patients' preference for either equipment examination procedures are discussed.

摘要

为了评估治疗状态以及鼻腔减充血状态下的鼻阻力,对53例因慢性或复发性鼻阻塞主诉前来就诊的非选择性连续患者,依次使用了两台定制的鼻阻力计(瑞典Cintec公司生产的Rhino-Comp和德国Atmos公司生产的鼻阻力计200),采用相同的主动前鼻测压法。将两台设备的最终平均结果进行配对统计分析。两台鼻阻力计提供的R150结果之间未发现相关性,尽管在比较两台设备所有R150结果的平均均值时,鼻阻力计200/Rhino-Comp的比值存在1.26至1.30的差异,这表明差异可能是由于各仪器固有的技术特性所致,且有产生系统性R150差异的趋势。当使用布罗姆斯系统参数V2值比较两台设备曲线的图形时,在呼气(而非吸气)结果中,Rhino-Comp和鼻阻力计200之间存在显著相关性,这表明两台设备产生的呼气曲线在图形上相似,至少在布罗姆斯系统半径2截距之前是这样。还发现有足够的数据支持这样一种可能性,即鼻阻力计200设备使用鼻适配器而非胶带进行前鼻孔阻塞的制造商选择可能会干扰鼻前庭功能,从而人为地影响阻力测量。最后,讨论了患者对两种设备检查程序的偏好。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验