Thomasson C G, Blijenberg B G, Leijnse B
J Clin Chem Clin Biochem. 1982 May;20(5):299-303. doi: 10.1515/cclm.1982.20.5.299.
We have evaluated the Technicon bound uricase method for the determination of uric acid in urine with the AutoAnalyzer II. The general analytical characteristics of the method, and the effect of urine on the immobilized uricase nylon tube reactor were investigated. The method was found to be a linear up to 7.0 mmol/l with aqueous standards and up to 5.0 mmol/l with urine samples. Between-days imprecision had a coefficient of variation (CV) of 1.6% for values of about 1.2 mmol/l, and 0.8% for values above 3.5 mmol/l. Within-run imprecision gave a CV of 0.7% for values of 1.07 and 2.95 mmol/l and a CV of 0.4% for a value of 4.77 mmol/l. The mean analytic recovery was 99.1% (range 93.8-103.0%). The sample interaction was 0.9%. The correlation with the phosphotungstate method and the manual Dutch standard method was good, but the enzymatic values were 20% lower than the phosphotungstate values. Storing the immobilized uricase nylon tube reactor at 4 degrees C, when not in use, prolonged the lifetime by nearly 50%. Urine samples were not different from aqueous uric acid standards in their effect on the stability of the uricase nylon tube reactor.
我们使用自动分析仪II评估了Technicon固定化尿酸酶法测定尿液中尿酸的情况。研究了该方法的一般分析特性以及尿液对固定化尿酸酶尼龙管反应器的影响。结果发现,该方法对于水溶液标准品在高达7.0 mmol/L时呈线性,对于尿液样品在高达5.0 mmol/L时呈线性。日间不精密度方面,约1.2 mmol/L的值变异系数(CV)为1.6%,高于3.5 mmol/L的值变异系数为0.8%。批内不精密度方面,1.07和2.95 mmol/L的值CV为0.7%,4.77 mmol/L的值CV为0.4%。平均分析回收率为99.1%(范围93.8 - 103.0%)。样品相互作用为0.9%。与磷钨酸盐法和荷兰手动标准法的相关性良好,但酶法测定值比磷钨酸盐法测定值低20%。固定化尿酸酶尼龙管反应器在不使用时于4℃储存,可使使用寿命延长近50%。尿液样品对尿酸酶尼龙管反应器稳定性的影响与水溶液尿酸标准品无异。