Oosterhuis B, van Boxtel C J
J Chromatogr. 1982 Nov 12;232(2):327-34. doi: 10.1016/s0378-4347(00)84172-9.
The sensitivity of electrochemical detection was combined with the selectivity of a bimodal high-performance liquid chromatographic system for the successful determination of salbutamol in human plasma. Following initial sample clean-up using Sep-Pak cartridges, analytes were passed first through a cation-exchange column, and then, after column switching, through a reversed-phase column. An amperometric detector with a rotated disc working electrode was used for detection. The detection limit was 0.5 ng/ml when 1.0 ml of plasma was used. The coefficient of variation was 9.8% at an average concentration of 4.7 ng/ml. The method was adequate for pharmacokinetic studies and for clinical applications.