Tronc D, Noël A
Strahlentherapie. 1978 Nov;154(11):780-5.
After a brief description of the Therac 20 Saturne linear accelerator a complete set of absorbed-dose distribution values is given. These values define the depths on the axis as a function of the depth dose and define the penumbra (as characterized by the positions of the intersections of the isodose curves with planes parallel to the phantom surface) for beams of X-rays and for beams of electrons. Tissue-maximum ratios are given for beams of X-rays. Analytical values for the electron depth dose curve are compared with the values obtained on the Sagittaire linear accelerator.
在对Therac 20 Saturne直线加速器进行简要描述之后,给出了一整套吸收剂量分布值。这些值定义了轴向上的深度作为深度剂量的函数,并定义了X射线束和电子束的半值层(以等剂量曲线与平行于体模表面的平面的交点位置为特征)。给出了X射线束的组织最大剂量比。将电子深度剂量曲线的分析值与在Sagittaire直线加速器上获得的值进行了比较。