Suppr超能文献

Absorbed dose distribution for X-ray beams and beams of electrons from the Therac 10 Neptune linear accelerator.

作者信息

Tronc D, Gayet P

出版信息

Strahlentherapie. 1980 Feb;156(2):125-8.

PMID:7355413
Abstract

After a brief presentation of the Therac 10 Neptune linear accelerator a complete set of dose distribution numerical values is given. These values define the depths on the axis as a function of the depth dose and define the penumbra (as characterized by the positions of the isodose curve intersections with parallel planes to the phantom surface) for beams of X-rays and for beams of electrons. Measurements of residual X-rays are given for a 10 MeV beam of electrons.