来自萨图恩直线加速器的6至20兆电子伏扫描电子束的辐射参数。

Radiation parameters of 6 to 20 MeV scanning electron beams from the Saturne linear accelerator.

作者信息

Pfalzner P M, Clarke H C

出版信息

Med Phys. 1982 Jan-Feb;9(1):117-20. doi: 10.1118/1.595230.

Abstract

Depth doses of the scanning electron beams from the Saturne Therac-20 linear accelerator at nominal energies of 6,9,13,17, and 20 MeV were measured in polystyrene using a thin window parallel plate ionization chamber. Central axis depth dose curves are derived and are analyzed according to the method of Brahme and Svensson. For each of the five electron energies, values are obtained for the most probable energy at the absorber surface Ep,0, the practical range Rp, the 50% range R50, the therapeutic range R85, the electron dose gradients, total collision energy losses, and other radiation parameters, and these are compared to corresponding values for electron beams from a 22 MeV medical microtron and a 20 MeV betatron.

摘要

使用薄窗平行板电离室在聚苯乙烯中测量了 Saturne Therac - 20 直线加速器在标称能量 6、9、13、17 和 20 MeV 下扫描电子束的深度剂量。得出中心轴深度剂量曲线,并根据布拉赫姆和斯文森的方法进行分析。对于这五种电子能量中的每一种,都获得了吸收体表面最可能能量 Ep,0、实际射程 Rp、50% 射程 R50、治疗射程 R85、电子剂量梯度、总碰撞能量损失以及其他辐射参数的值,并将这些值与来自 22 MeV 医用电子回旋加速器和 20 MeV 电子感应加速器的电子束的相应值进行比较。