[角膜水肿中压平眼压计与指测眼压的差异]

[Disparity between applanation and impression tonometry in corneal edema].

作者信息

Kulnig W

出版信息

Klin Monbl Augenheilkd. 1982 Nov;181(5):370-2. doi: 10.1055/s-2008-1055246.

Abstract

In acute glaucoma with corneal edema the applanationtonometer shows significantly lower values than the Schiötz instrument. In an experiment with rabbit corneae characteristic alterations in the edematous cornea were found, the physical interpretation of which, together with the tonometer specifications, were regarded as the cause of the disparity. The consequences go beyond pure academic interest if high intraocular pressure occurs concomitantly with insufficient retinal vascular function.

摘要

在伴有角膜水肿的急性青光眼患者中,压平眼压计显示的值明显低于 Schiötz 眼压计。在一项对兔角膜的实验中,发现水肿角膜有特征性改变,其物理解释连同眼压计的规格被视为差异的原因。如果高眼压与视网膜血管功能不足同时出现,其后果就不仅仅是纯学术兴趣范畴了。