Popli S, Ing T S, Daugirdas J T, Kheirbek A O, Viol G W, Vilbar R M, Gandhi V C
Artif Organs. 1982 Aug;6(3):312-5. doi: 10.1111/j.1525-1594.1982.tb01679.x.
Five severe reactions occurred in four maintenance hemodialysis patients 1 to 5 minutes after initiating dialysis with Cuprophan capillary dialyzers. All reactions were life-threatening and one resulted in death. Inadequate rinsing of the dialyzers was probably the cause of the reactions. The severe reactions were managed by immediate discontinuation of dialysis and the institution of supportive treatment. Antianaphylactic measures were also attempted, but their therapeutic effectiveness remains to be determined.
4名维持性血液透析患者在使用铜仿膜毛细血管透析器开始透析1至5分钟后发生了5起严重反应。所有反应均危及生命,其中1例导致死亡。透析器冲洗不充分可能是反应的原因。通过立即停止透析并采取支持性治疗措施来处理严重反应。还尝试了抗过敏措施,但其治疗效果尚待确定。