Sherrard E S, Buckley R J
Invest Ophthalmol Vis Sci. 1981 Jun;20(6):816-20.
The value of clinical specular microscopy of the corneal endothelium has been limited by the extremely small field of view of the image obtained. This is largely a result of an overlying brilliant reflection from the objective (or air)/cornea interface. A modification to the applanating objective cone of the contract specular microscope is described, which virtually eliminates this reflection and gives a visible field of almost 1 mm2 of the endothelium. The theory behind the objective modification is presented, and other advantages of the system, such as reduction of eye movement, are noted.
角膜内皮临床镜面显微镜检查的价值一直受到所获图像视野极小的限制。这主要是由于物镜(或空气)/角膜界面的强烈反射所致。本文描述了对接触式镜面显微镜的压平物镜锥体的一种改进方法,该方法几乎消除了这种反射,并给出了近1平方毫米内皮细胞的可见视野。文中介绍了物镜改进背后的理论,并指出了该系统的其他优点,如减少眼球运动。