Suppr超能文献

Permanent pacemaker malfunction: diagnostic aspects of temporary pacing.

作者信息

Berman N D

出版信息

Can Med Assoc J. 1980 Aug 9;123(3):190-3.

Abstract

A temporary pacing electrode can function as a diagnostic as well as a therapeutic tool. This is illustrated in two patients whose permanent pacemakers unexpectedly failed. In the first patient a demand pacemaker was inhibited by a magnet rather than converting to the asynchronous mode. In the second the pacemaker appeared to be producing low-voltage potentials not detectable on the surface electrocardiogram. The presence of a temporary pacing electrode can be useful for defining the cause of pacemaker failure and the nature of any associated arrhythmias.

摘要
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/949a/1704732/9afeaa2b2732/canmedaj01463-0030-a.jpg

相似文献

2
Temporary pacemaking by general internists.
Arch Intern Med. 1983 Aug;143(8):1531-3.
5
Runaway pacemaker. Unpredictable pacemaker failure.
Arch Intern Med. 1979 Oct;139(10):1190-1.
8
[Rhythmic pitfalls in cardiac pacing].
Arch Mal Coeur Vaiss. 1998 Mar;91 Spec No 1:61-9.

本文引用的文献

1
Suppression of demand pacemakers by inactive pacemaker electrodes.
Circulation. 1972 Feb;45(2):319-23. doi: 10.1161/01.cir.45.2.319.
2
Demand pacemaker malfunction due to abnormal sensing. Report of two cases.
Circulation. 1974 Aug;50(2):389-94. doi: 10.1161/01.cir.50.2.389.
3
Reciprocal beats initiated by artificial pacemaker.
Heart Lung. 1976 Jan-Feb;5(1):124-6.
4
A new hazard in the use of an external demand pacemaker.
J Electrocardiol. 1978 Jan;11(1):93-6. doi: 10.1016/s0022-0736(78)80038-7.
5
Apparent malfunction of demand pacemakers due to spurious potentials generated by contact between two endocardial electrodes.
Pacing Clin Electrophysiol. 1978 Oct;1(4):531-4. doi: 10.1111/j.1540-8159.1978.tb03517.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验