[Quality control of computer-assisted arrhythmia monitoring].

作者信息

Ritz R

出版信息

Schweiz Med Wochenschr. 1981 Jun 27;111(26):997-1000.

PMID:7268355
Abstract

In the management of coronary care patients, continuous ECG analysis is best performed by a computerized monitoring system. For more than 2 years all the patients in this CCU have been monitored by the 78220 Hewlett-Packard arrhythmia monitoring system. As a quality control of the system the data of 15,068 alarm situations in 300 patients were first analyzed: the specificity of the system varied from 3% to 83% according to type and severity of arrhythmias. To evaluate the sensitivity of the system the amount of VES registered by Holter monitoring was then compared with the analysis by computerized monitoring during the same period of time: in a total of over 10,000 VES in 14 patients the sensitivity was 90%. Although the personnel of a CCU still have to deal with a large number of false positive alarms, a computerized arrhythmia monitoring system offers a higher degree of security for the patient than conventional monitoring.

摘要

在冠心病监护患者的管理中,连续心电图分析最好由计算机化监测系统进行。在两年多的时间里,该冠心病监护病房的所有患者均由惠普78220心律失常监测系统进行监测。作为对该系统的质量控制,首先分析了300例患者中15068次警报情况的数据:根据心律失常的类型和严重程度,该系统的特异性从3%到83%不等。为了评估该系统的敏感性,随后将动态心电图监测记录的室性早搏数量与同期计算机化监测的分析结果进行了比较:在14例患者共10000多次室性早搏中,敏感性为90%。尽管冠心病监护病房的工作人员仍需处理大量的假阳性警报,但与传统监测相比,计算机化心律失常监测系统为患者提供了更高程度的安全保障。