Abstract

A sensitive method (Clin. Chem. 26: 327--331, 1980) for serum iron, in which the color reagent Ferrozine is used, is modified and adapted to the Abbott ABA-100 discrete analyzer. The standard curve is linear to at least 10 mg/L and the method showed day-to-day precision (CV) of 2.4% for a 1.03 mg/L sample (n = 63) and 1.9% for a 2.13 mg/L sample (n = 63). Lower values were obtained than with the modified continuous-flow technique of Giovanniello et al., but the correlation was good (r = 0.98). Bilirubin and copper do not interfere; hemoglobin and gross lipemia interfere only slightly. The total iron-binding capacity, based on Ramsay's method, was evaluated with regard to the effect of adding various amounts of magnesium carbonate. Results led us to use a ratio of approximately 180 mg of magnesium carbonate to each 5 micrograms of excess iron added. Day-to-day, the method for total iron-binding capacity gave a CV of 3.1% for a 2.55 mg/L sample, 2.8% for a 3.63 mg/L sample.

摘要

一种使用显色剂菲洛嗪测定血清铁的灵敏方法(《临床化学》26: 327 - 331, 1980)经过改进,适用于雅培ABA - 100离散分析仪。标准曲线至少在10 mg/L范围内呈线性,该方法对于1.03 mg/L的样本(n = 63)显示出2.4%的日间精密度(CV),对于2.13 mg/L的样本(n = 63)显示出1.9%的日间精密度。与乔瓦尼罗等人改进的连续流动技术相比,得到的值较低,但相关性良好(r = 0.98)。胆红素和铜不产生干扰;血红蛋白和严重脂血仅产生轻微干扰。基于拉姆齐方法评估了总铁结合力,考虑了添加不同量碳酸镁的影响。结果使我们采用每添加5微克过量铁对应约180毫克碳酸镁的比例。对于总铁结合力的方法,对于2.55 mg/L的样本,日间CV为3.1%,对于3.63 mg/L的样本,日间CV为2.8%。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验