Baxter B S, Sorenson J A
Invest Radiol. 1981 Jul-Aug;16(4):337-41. doi: 10.1097/00004424-198107000-00017.
Computed tomography scanners may be used to make anatomic size measurement; however, manipulating the viewer controls (especially the window center) can have a significant influence on the apparent size of structures in the image. A special phantom was constructed to study the effect of window center adjustments on the apparent size of lone cylindrical objects (no variation within slice thickness) and also for spherical objects (maximum variation within slice thickness). Using this phantom, the authors were able to identify several commonly encountered situations in which both the apparent size and CT numbers may be seriously in error. It was found that for cylindrical objects larger than about one transverse resolution element and aligned with the scanner axis, CT numbers may be determined accurately; however, the apparent diameter changes by several millimeters as the window center setting is changed. CT numbers for spheres are inaccurate when the diameter is comparable to or less than the slice thickness and, as a result, diameters of spheres cannot be measured accurately with fixed window center settings. Diameters of spheres can be measured accurately (+/- 1 mm) if the sphere is centered in the slice and the full width at half maximum of the CT number profile is used.
计算机断层扫描扫描仪可用于进行解剖尺寸测量;然而,操作观察器控件(尤其是窗中心)会对图像中结构的表观尺寸产生重大影响。构建了一个特殊的体模,以研究窗中心调整对单个圆柱形物体(层厚内无变化)以及球形物体(层厚内变化最大)表观尺寸的影响。使用该体模,作者能够识别出几种常见情况,在这些情况下,表观尺寸和CT值都可能出现严重误差。结果发现,对于大于约一个横向分辨率单元且与扫描仪轴对齐的圆柱形物体,CT值可以准确测定;然而,随着窗中心设置的改变,表观直径会变化几毫米。当球体直径与层厚相当或小于层厚时,球体的CT值不准确,因此,使用固定的窗中心设置无法准确测量球体的直径。如果球体位于切片中心并使用CT值轮廓的半高全宽,则可以准确测量球体的直径(±1毫米)。