Suppr超能文献

头部CT扫描期间内部散射对辐射剂量的贡献。

The contribution of internal scatter to radiation dose during CT scan of the head.

作者信息

Chopp M, Ewald L, Hartson M

出版信息

Neuroradiology. 1981;22(3):145-50. doi: 10.1007/BF00346757.

Abstract

An acrylic head phantom was irradiated during a computed tomographic scan with four commercial scanners. Measurements of the spatial distribution of the radiation dose on the surface and internal to the phantom were performed for the scan plane and the scattered beam at various distances from the scan plane. The surface scatter dose was found to be considerably smaller than that for internal scatter. A significant increase in radiation exposure within the head phantom due to internal radiation scatter, and an asymmetrical primary beam profile for dual slice scanners were also noted.

摘要

在使用四台商用扫描仪进行计算机断层扫描期间,对一个丙烯酸头部模型进行了照射。针对扫描平面以及距扫描平面不同距离处的散射束,测量了模型表面和内部辐射剂量的空间分布。结果发现,表面散射剂量远小于内部散射剂量。还注意到,由于内部辐射散射,头部模型内的辐射暴露显著增加,并且双切片扫描仪的初级光束轮廓不对称。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验