Aldrich J E, Andrew J W
J Can Assoc Radiol. 1981 Dec;32(4):202-5.
Thermoluminescent dosimeters (TLD) have been sent by mail to all hospitals in Nova Scotia to survey the radiographic and fluoroscopic x-ray exposures to patients, For each radiographic system a TLD was placed in the x-ray field during a routine abdominal radiographic procedure on an average size patient (20 cm AP). The TLD was not visible on the film and in no way effected the examination. Fluoroscopic exposure rates were measured using TLD and a 20 cm water phantom. The dosimeter was exposed for one minute using technique factors for an abdomen of average size. The average abdominal exposure to a 20 cm thick patient on 137 different radiographic units was .74 R. The average fluoroscopic exposure rat to a 20 cm water phantom using 54 different fluoroscopic systems was 5.4 R/min.
热释光剂量计(TLD)已通过邮件发送给新斯科舍省的所有医院,以调查患者接受的X射线摄影和荧光透视X射线照射情况。对于每个X射线摄影系统,在对平均体型患者(前后径20厘米)进行常规腹部X射线摄影检查时,将一个TLD放置在X射线照射野中。TLD在胶片上不可见,且绝不会影响检查。使用TLD和一个20厘米水模体测量荧光透视照射率。剂量计使用平均体型腹部的技术参数进行一分钟的照射。在137个不同的X射线摄影设备上,平均体型为20厘米厚的患者接受的腹部照射量为0.74伦琴。使用54个不同的荧光透视系统对20厘米水模体进行照射,平均荧光透视照射率为5.4伦琴/分钟。