Enzyme electrode for phenol.

作者信息

Kjellén K G, Neujahr H Y

出版信息

Biotechnol Bioeng. 1980 Feb;22(2):299-310. doi: 10.1002/bit.260220205.

Abstract

An enzyme electrode is described for quantitative determination of phenol at micromolar concentrations. Immobilized phenol hydroxylase is attached to the surface of a Clark oxygen electrode. The maximum rate of oxygen consumption is linearly dependent on phenol concentration over the 0.5-50 microM range. The electrode can be used for at least 150 assays without an activity loss. Readout is very rapid--within 30 sec of sample addition. The electrode response is independent of pH between pH 6.5 and 9.5. The response increases linearly with temperature in the interval 10-40 degrees C. It is necessary to incubate the enzyme electrode in a buffer containing NADPH for a few minutes before the addition of sample. This is to make the electrode response independent of the diffusion rate of this cosubstrate. This and other diffusional effects on the performance of the phenol electrode are discussed.

摘要

本文描述了一种用于定量测定微摩尔浓度苯酚的酶电极。固定化的苯酚羟化酶附着在Clark氧电极表面。在0.5 - 50微摩尔范围内,最大耗氧速率与苯酚浓度呈线性相关。该电极可用于至少150次测定而不会损失活性。读数非常迅速——加入样品后30秒内即可完成。电极响应在pH 6.5至9.5之间与pH无关。在10 - 40摄氏度的区间内,响应随温度呈线性增加。在加入样品之前,需要将酶电极在含有NADPH的缓冲液中孵育几分钟。这是为了使电极响应与这种辅酶的扩散速率无关。本文还讨论了这一点以及其他扩散效应对苯酚电极性能的影响。