关于使用18兆电子伏特电子直线加速器照射患者的辐射防护测量。

Measurements on the radiation protection of patients irradiated with an 18-MeV electron linear accelerator.

作者信息

Nüsslin F, Hassenstein E

出版信息

Strahlentherapie. 1980 Feb;156(2):120-4.

PMID:7355412
Abstract

In order to quantify the influence of some physical and geometrical treatment parameters on the stray radiation dose absorbed by patients undergoing radiotherapy on an 18-MeV linear accelerator, dose distributions were measured in a polystyrene phantom by means of LiF-dosimeters. For 10-MV X-rays, 6-MeV and 18-MeV electron radiation dose profiles, depth dose and isodose curves were obtained at various field sizes and treatment couch positions. By covering the phantom with 1 or 2 mm thick lead sheets a remarkable dose reduction was found. A two-exponential expression was derived to estimate stray radiation dosage in routine radiotherapy. The validity of the calculation method was demonstrated by comparing the gonadal load measured in an Alderson phantom with the calculated data.

摘要

为了量化一些物理和几何治疗参数对在18兆电子伏直线加速器上接受放射治疗的患者所吸收的杂散辐射剂量的影响,通过LiF剂量计在聚苯乙烯模体中测量了剂量分布。对于10兆伏X射线、6兆伏和18兆伏电子辐射,在不同野大小和治疗床位置获得了剂量分布、深度剂量和等剂量曲线。通过用1或2毫米厚的铅板覆盖模体,发现剂量显著降低。推导了一个双指数表达式来估计常规放射治疗中的杂散辐射剂量。通过将在Alderson模体中测量的性腺负荷与计算数据进行比较,证明了计算方法的有效性。