使用以管状形式共固定化的己糖激酶和葡萄糖-6-磷酸脱氢酶对血清中的葡萄糖进行连续流动分析。

Continuous-flow analysis for glucose in serum, with use of hexokinase and glucose-6-phosphate dehydrogenase co-immobilized in tubular form.

作者信息

León L P, Chu D K, Snyder L R, Horváth C

出版信息

Clin Chem. 1980 Jan;26(1):123-9.

PMID:7356545
Abstract

Hexokinase and glucose-6-phosphate dehydrogenase, co-immobilized at the inner wall of plastic tubing, were used with continuous-flow analyzers to measure serum glucose. The performance of the enzyme-containing coils meets the hydraulic requirements of the continuous-flow instruments, and the average carryover does not exceed 2.2%. A glucose concentration change of 1 g/L yields a 0.08 change in absorbance. There is no significant deviation from linearity for glucose concentrations up to 5 g/L. The method is similar to the Reference Method proposed by the Food and Drug Administration, and results by the two methods agree and correlate well (r = 0.999). The operational stability of the enzyme tubes is sufficient to allow the analysis of at least 10 000 serum samples. We investigated in detail the effect of reagent concentrations and temperature on the performance of the enzyme coils, and the results shed light on the nature of the heterogeneous dual-enzymic reaction.

摘要

将己糖激酶和葡萄糖-6-磷酸脱氢酶共固定在塑料管材内壁上,与连续流动分析仪一起用于测量血清葡萄糖。含酶盘管的性能满足连续流动仪器的水力要求,平均残留率不超过2.2%。葡萄糖浓度每变化1 g/L,吸光度变化0.08。葡萄糖浓度高达5 g/L时,与线性关系无显著偏差。该方法与美国食品药品监督管理局提出的参考方法相似,两种方法的结果一致且相关性良好(r = 0.999)。酶管的操作稳定性足以允许分析至少10000份血清样本。我们详细研究了试剂浓度和温度对酶盘管性能的影响,结果揭示了非均相双酶反应的本质。