Suppr超能文献

早产儿的热平衡:对流加热暖箱和辐射保暖台的比较效果,有无塑料隔热罩。

Heat balance in premature infants: comparative effects of convectively heated incubator and radiant warmer, with without plastic heat shield.

作者信息

Bell E F, Weinstein M R, Oh W

出版信息

J Pediatr. 1980 Mar;96(3 Pt 1):460-5. doi: 10.1016/s0022-3476(80)80698-6.

Abstract

Insensible water loss, oxygen consumption, and carbon dioxide production were measured in eight premature infants under four different conditions: in conventional single-walled incubator with and without plastic heat shield, and under radiant warmer with and without heat shield. IWL was greater under the radiant warmer (3.40 +/- 1.50 ml/kg/hour, mean +/- SD) than in the incubator (2.37 +/- 1.15 ml/kg/hour) when both were compared without heat shield. Addition of the heat shield reduced IWL in the incubator (2.13 +/- 0.76 ml/kg/hour) but not under the radiant warmer (3.37 +/- 0.94 ml/kg/hour). There were no significant differences in VO2 or respiratory quotient between any two of the four study conditions.

摘要

在四种不同条件下,对八名早产儿的不显性失水量、耗氧量和二氧化碳生成量进行了测量:分别在有和没有塑料隔热罩的传统单壁暖箱中,以及在有和没有隔热罩的辐射保暖台上。当两者都没有隔热罩时,辐射保暖台下的不显性失水量(3.40±1.50毫升/千克/小时,平均值±标准差)高于暖箱中的(2.37±1.15毫升/千克/小时)。添加隔热罩可降低暖箱中的不显性失水量(2.13±0.76毫升/千克/小时),但在辐射保暖台下则没有降低(3.37±0.94毫升/千克/小时)。四种研究条件中的任意两种之间,耗氧量或呼吸商均无显著差异。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验