Bell E F, Weinstein M R, Oh W
J Pediatr. 1980 Mar;96(3 Pt 1):460-5. doi: 10.1016/s0022-3476(80)80698-6.
Insensible water loss, oxygen consumption, and carbon dioxide production were measured in eight premature infants under four different conditions: in conventional single-walled incubator with and without plastic heat shield, and under radiant warmer with and without heat shield. IWL was greater under the radiant warmer (3.40 +/- 1.50 ml/kg/hour, mean +/- SD) than in the incubator (2.37 +/- 1.15 ml/kg/hour) when both were compared without heat shield. Addition of the heat shield reduced IWL in the incubator (2.13 +/- 0.76 ml/kg/hour) but not under the radiant warmer (3.37 +/- 0.94 ml/kg/hour). There were no significant differences in VO2 or respiratory quotient between any two of the four study conditions.