Suppr超能文献

使用一对小型圆盘热敏电阻对脑脊液分流管通畅性进行热敏测定。

Thermosensitive determination of CSF shunt patency with a pair of small disc thermistors.

作者信息

Chiba Y, Yuda K

出版信息

J Neurosurg. 1980 May;52(5):700-4. doi: 10.3171/jns.1980.52.5.0700.

Abstract

A simple method is presented for detecting the patency of ventriculoperitoneal shunts with a pair of small disc thermistors. In an experimental model, the equipment detected definite temperature changes along the shunt tube. The extent of the temperature difference was proportional to the flow velocities, ranging from 5 to 60 ml/hr (120 to 1440 ml/24 hrs). A clinical study consisted of 26 trials in 23 hydrocephalic children and 10 trials in nine adult patients with normal-pressure hydrocephalus. Three of the 26 trials in the children were unsuccessful because of patient irritability and lack of cooperation. All trials for adult patients were successful. In all 25 patients, whose clinical pictures suggested functioning shunt, accurate confirmation of shunt patency was obtained. In six patients whose clinical pictures suggested questionable shunt function, this method revealed that all shunts were patent, and computerized tomography and follow-up clinical data supported this finding. In two patients whose clinical pictures suggested shunt malfunction, there was no indicator deflection, and shunt obstruction was proved at operation. This method is effective in checking shunt function, is simple, and may be repeated at frequent intervals.

摘要

介绍了一种用一对小型圆盘热敏电阻检测脑室腹腔分流管通畅性的简单方法。在一个实验模型中,该设备检测到沿分流管有明确的温度变化。温差程度与流速成正比,流速范围为5至60毫升/小时(120至1440毫升/24小时)。一项临床研究包括对23名脑积水儿童进行的26次试验以及对9名正常压力脑积水成年患者进行的10次试验。儿童的26次试验中有3次因患者烦躁和不配合而未成功。成年患者的所有试验均成功。在所有25例临床症状提示分流管功能正常的患者中,均准确证实了分流管通畅。在6例临床症状提示分流管功能存疑的患者中,该方法显示所有分流管均通畅,计算机断层扫描和后续临床数据支持这一发现。在2例临床症状提示分流管故障的患者中,没有指标偏转,手术证实存在分流管梗阻。该方法在检查分流管功能方面有效、简单,且可频繁重复。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验