Suppr超能文献

自动超阈值静态视野检查法。

Automated suprathreshold static perimetry.

作者信息

Johnson C A, Keltner J L

出版信息

Am J Ophthalmol. 1980 May;89(5):731-41. doi: 10.1016/0002-9394(80)90296-2.

Abstract

Automated (Fieldmaster) and manual kinetic perimetry were performed in 1,019 eyes with various visual disorders. tunder appropriate test conditions, automated Fieldmaster evaluations showed a 96.1% detection rate and a 4.7% false alarm rate for detection of visual field defects. Preliminary results for 168 eyes showed that sequential follow-up testing with the Fieldmaster correctly identified progression or regression of visual field loss in more than 94% (34 of 36) of the eyes with visual field changes. The Fieldmaster correctly determined no change in visual field characteristics for 99% (131 of 132) of the eyes with stable visual fields. These data indicate that the Fieldmaster is an excellent device for detecting visual field defects, and also has some capabilities for monitoring progression or regression of visual field loss.

摘要

对1019只患有各种视觉障碍的眼睛进行了自动(Fieldmaster)和手动动态视野检查。在适当的测试条件下,自动Fieldmaster评估显示,检测视野缺损的检出率为96.1%,误报率为4.7%。168只眼睛的初步结果表明,使用Fieldmaster进行连续随访测试能正确识别出超过94%(36只中的34只)视野发生变化的眼睛的视野缺损进展或消退情况。对于99%(132只中的131只)视野稳定的眼睛,Fieldmaster正确判定其视野特征没有变化。这些数据表明,Fieldmaster是检测视野缺损的优秀设备,并且还具备监测视野缺损进展或消退的一些能力。