Hong C, Kitazawa Y, Shirato S
Int Ophthalmol. 1982 Feb;4(3):151-7. doi: 10.1007/BF00161905.
The efficacy of automated suprathreshold perimetry with Fieldmaster automated perimeter (model 101 PR and 200) was evaluated in the screening for very early glaucomatous visual field defects which failed to be detected with kinetic perimetry using a Goldman perimeter. Evaluation of 117 eyes with elevated intraocular pressure revealed a small, relative defect in the central field in 14 eyes which was confirmed by static perimetry with a Tübinger perimeter. Although Fieldmaster model 200 is equipped with more target positions than model 101 PR, detectability of early defects was almost identical between two models. Automated suprathreshold perimetry with either model of Fieldmaster perimeter is useful in detecting the early changes which may be missed by kinetic perimetry with a Goldman perimeter. Based on the difference in the false positive rate (45.6% with model 200 vs 26.7% with model 101 PR) the screening with 101 PR seems to be more rewarding.
使用Fieldmaster自动视野计(型号101 PR和200)进行自动超阈值视野检查,在筛查极早期青光眼性视野缺损方面的功效进行了评估,这些缺损用戈德曼视野计进行动态视野检查未能检测到。对117只眼压升高的眼睛进行评估,发现14只眼睛中央视野有小的相对缺损,经 Tübinger视野计静态视野检查得到证实。尽管Fieldmaster 200型比101 PR型配备了更多的目标位置,但两种型号对早期缺损的可检测性几乎相同。使用任何一种型号的Fieldmaster视野计进行自动超阈值视野检查,对于检测动态戈德曼视野检查可能遗漏的早期变化都很有用。基于假阳性率的差异(200型为45.6%,101 PR型为26.7%),使用101 PR型进行筛查似乎更有价值。