用于测定治疗期间血浆或血清中胺碘酮浓度的快速高效液相色谱法。

Rapid high-performance liquid chromatographic method for the measurement of amiodarone in blood plasma or serum at the concentrations attained during therapy.

作者信息

Flanagan R J, Storey G C, Holt D W

出版信息

J Chromatogr. 1980 Jan 18;187(2):391-8. doi: 10.1016/s0021-9673(00)80471-9.

Abstract

A simple high-performance liquid chromatographic method has been developed for the measurement of the antiarrhythmic drug amiodarone in small (200 microliter) volumes of plasma or serum. After addition of 2 mole/l phosphate solution, pH 4.5 (20 microliter), containing the internal standard, the sample is vortex-mixed with diisopropyl ether (200 microliter) for 30 sec. A portion (100 microliter) of the resulting extract is analysed on a microparticulate (5 micron) silica column using methanol-diethyl ether (85:15) containing perchloric acid (0.02% v/v) as the mobile phase, and the absorption of the column effluent is monitored at 240 nm. No endogenous sources of interference have been observed, and interference from other drugs is minimal. The procedure is rapid, an analysis in duplicate taking less than 15 min to complete. The limit of sensitivity of the assay is 0.05 mg/l, and the concentrations of amiodarone measured in plasma samples from patients under treatment with this compound ranged from 0.15 to 4.5 mg/l.

摘要

已开发出一种简单的高效液相色谱法,用于测定小体积(200微升)血浆或血清中的抗心律失常药物胺碘酮。加入含内标的2摩尔/升pH 4.5的磷酸盐溶液(20微升)后,将样品与二异丙醚(200微升)涡旋混合30秒。取所得提取物的一部分(100微升),在微粒(5微米)硅胶柱上进行分析,以含高氯酸(0.02% v/v)的甲醇-乙醚(85:15)作为流动相,并在240纳米处监测柱流出物的吸光度。未观察到内源性干扰源,其他药物的干扰也极小。该方法快速,重复分析不到15分钟即可完成。该测定法的灵敏度极限为0.05毫克/升,在用该化合物治疗的患者血浆样本中测得的胺碘酮浓度范围为0.15至4.5毫克/升。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验