Rucki R J, Ross A, Moros S A
J Chromatogr. 1980 Apr 4;190(2):359-65. doi: 10.1016/s0021-9673(00)88240-0.
An amperometric flow-through detector with a carbon paste working electrode was utilized as a high-performance liquid chromatographic (HPLC) detector to determine procarbazine hydrochloride, an antineoplastic agent, in both buffer solution and biological fluids. The HPLC system included an amino-cyano stationary phase and an aqueous (pH 7)-methanolic mobile phase which enabled the separation of procarbazine from its only electroactive degradation product, N-isopropyl-alpha-(2-methylhydrazono)-p-toluamide. The electrochemical detector, with an approximate limit of detection of 2 ng procarbazine injected, was 20 times more sensitive to procarbazine than a typical UV detector. The low dead volume (1 microliter) and superior selectivity of the electrochemical detector enabled the HPLC determination of procarbazine in untreated human urine and plasma.