Suppr超能文献

可变文丘里型氧气面罩的性能分析。

An analysis of the performance of a variable venturi-type oxygen mask.

作者信息

Woolner D F, Larkin J

出版信息

Anaesth Intensive Care. 1980 Feb;8(1):44-51. doi: 10.1177/0310057X8000800109.

Abstract

The theoretical performance of the Hudson Multivent mask is considered. A method is described of determining the flow-weighted mean inspired oxygen concentration produced by the mask. Using this method, it was found that the inspired oxygen concentration was predictable only at the 24% setting using the recommended flow rates, but that increasing the supplied-oxygen flow rates above the recommended levels resulted in the return of predictable function. At settings above 30% predictable function was found to be unlikely, whatever the supplied oxygen flow rate. Recommendations are made regarding the use of this device.

摘要

对哈德逊多通气孔面罩的理论性能进行了研究。描述了一种确定该面罩产生的流量加权平均吸入氧浓度的方法。使用该方法发现,仅在使用推荐流速的24%设置下,吸入氧浓度是可预测的,但将供应氧流速提高到推荐水平以上会使功能恢复可预测性。发现在30%以上的设置下,无论供应氧流速如何,都不太可能实现可预测的功能。针对该设备的使用提出了建议。