We describe a kinetic procedure using the LKB reaction rate analyser for the calibration of an AutoAnalyzer alkaline phosphatase method. Measurements were made at zero incubation time with a special lid for the thermostated tunnel of the LKB reaction rate analyzer using assay conditions appropriate to the AutoAnalyzer method. The molar absorptivity of p-nitrophenol was estimated on the LKB reaction rate analyser used for assay of the enzyme standards. Other variables affecting the accuracy of the measurement were also determined.
我们描述了一种使用LKB反应速率分析仪校准自动分析仪碱性磷酸酶方法的动力学程序。在零孵育时间下,使用适合自动分析仪方法的分析条件,通过LKB反应速率分析仪恒温通道的特殊盖子进行测量。对用于酶标准品测定的LKB反应速率分析仪上对硝基苯酚的摩尔吸光系数进行了估算。还确定了影响测量准确性的其他变量。